MODEL 1600
362431
STANDARD THERMOCOUPLE MODEL 1600
STANDARD THERMOCOUPLE MODEL 1600
0.00

Product Description

The Isothermal range of Thermocouple Standards are the result of many years development. The type R and S standards will cover the range from 0°C to 1600°C.

The measuring assembly comprises a 7mm x 300mm or 600mm gas tight 99.7% recrystallized alumina sheath inside which is a 2.5mm diameter twin bore tube holding the thermocouple.

The inner 2.5mm assembly is removable since some calibration laboratories will only accept fine bore tubed thermocouples and some applications require fine bore tubing.

The covered noble metal thermocouple wire connects the measuring sheath to the reference sheath which is a 4.5mm x 250mm stainless steel sheath suitable for referencing in a 0°C reference system. Two thermo electrically free multistrand copper wires (teflon coated) connect the thermocouple to the voltage measuring device.

The thermocouple material is continuous from the hot or measuring junction to the cold, or referencing junction.

Tech Spec

TECHNICAL FEATURE VALUE
Model 1600
Hot Sheath Temperature Range 0°C to 1600°C (R or S)
Emf vs Temperature According to relevant document
Response Time 5 minutes
Hot Junction Dimensions see diagram
Connecting Cable See diagram
Cold Junction 250.0mm long x 4.5 diameter
Copper Extension Wires 2000mm
Immersion 100mm min.
Case Dimensions Height 65mm, Width 710mm, Depth 165mm, Gross Weight 900g
Feature Removeable inner assembly
The standard thermocouple described can be supplied in the following noble metal combinations: Type R: Platinum v Platinum 13% Rhodium or Type S: Platinum v Platinum 10% Rhodium
How to order Model 1600 type R/300, Model 1600 type R/600, Model 1600 type S/300, Model 1600 type S/600. If cold junction not required, specify NCJ. UKAS Calibration included.

Downloads

  • Açıklama
    • Product Description

      The Isothermal range of Thermocouple Standards are the result of many years development. The type R and S standards will cover the range from 0°C to 1600°C.

      The measuring assembly comprises a 7mm x 300mm or 600mm gas tight 99.7% recrystallized alumina sheath inside which is a 2.5mm diameter twin bore tube holding the thermocouple.

      The inner 2.5mm assembly is removable since some calibration laboratories will only accept fine bore tubed thermocouples and some applications require fine bore tubing.

      The covered noble metal thermocouple wire connects the measuring sheath to the reference sheath which is a 4.5mm x 250mm stainless steel sheath suitable for referencing in a 0°C reference system. Two thermo electrically free multistrand copper wires (teflon coated) connect the thermocouple to the voltage measuring device.

      The thermocouple material is continuous from the hot or measuring junction to the cold, or referencing junction.

      Tech Spec

      TECHNICAL FEATURE VALUE
      Model 1600
      Hot Sheath Temperature Range 0°C to 1600°C (R or S)
      Emf vs Temperature According to relevant document
      Response Time 5 minutes
      Hot Junction Dimensions see diagram
      Connecting Cable See diagram
      Cold Junction 250.0mm long x 4.5 diameter
      Copper Extension Wires 2000mm
      Immersion 100mm min.
      Case Dimensions Height 65mm, Width 710mm, Depth 165mm, Gross Weight 900g
      Feature Removeable inner assembly
      The standard thermocouple described can be supplied in the following noble metal combinations: Type R: Platinum v Platinum 13% Rhodium or Type S: Platinum v Platinum 10% Rhodium
      How to order Model 1600 type R/300, Model 1600 type R/600, Model 1600 type S/300, Model 1600 type S/600. If cold junction not required, specify NCJ. UKAS Calibration included.