Microtest 6365
362372
Microtest 6365 LCR Meter
Microtest 6365 LCR Meter
0.00

Product Description:

The 6370 LCR Meter that is provided with high accuracy and speed. It is a cost and multiple- function utility and is able to perform the component test entirely. The friendly operation interface design is in synchronization with the international trend and easy to operate to import into the production lines quickly. Because of the high-precision values of measured data and superior efficacy, it must be a necessary testing instrument of components and materials. It supports the interfaces of  RS232、Handler  and GPIB PC connections.

  • Basic accuracy up to 0.1%
  • LCR Meters are all compatible with convenient and efficient Meter Link Software
  • Overall Compact design, good for production line use
  • User-friendly interface, allows for production line personnel to operate efficiently

Product Specification:

Measurement Data

   
Test Frequency 100Hz-20KHz
Points Of Test Frequency 372 points
Basic Accuracy ± 0.1%
AC Drive Level 0.05V、0.1V、0.25V、0.5V、1V
DC Drive Level 0.5V
Output Impedance 100Ω
Frequency trimming resolution 10 Digits
Frequency output accuracy ± 0.01%
AC test speed (Fast) 5 times/sec
AC test speed (Medium) 3 times/sec
AC test speed (Slow) 2 times/sec
DC test speed (Fast) 5 times/sec
DC test speed  (Medium) 3 times/sec
DC test speed  (Slow) 2 times/sec
Bias voltage 2V

Specifications

   
ESR Series/Parallel
Mathematical Mode Absolute/Percentage
Test Range Automatic/Lock
Indicator Pass/Fail/Bias LED and Built-in speaker
Output Terminal BNC connector
Safety EN61010-1
Immunity EN61326-1
Emissions EN55022 Class A
Calibration Open Circuit / Short Circuit: Avoid the misreading caused by the stray capacitance from kelvin clips and fixtures
Power Supply Voltage 90-132Vac or 198-264Vac (selectable)
  Frequency 47-63Hz
Display 240*64 dot-matrix display
Interface RS-232, GPIB (option), Handler (option)
Flash Memory Storage of 127 rewritable data sets
Operation Auto、Manual、Remote Control、RS-232、GPIB (option)
Environment Temperature : 10°C to 40°C    Humidity : 20 ~ 90%
Dimension (W*H*D) 340x88x330 mm
Weight 5 Kg

Parameters Measurement Ranges

   
|Z| R X 0.1 mΩ – 100MΩ
|Y| G B 10nS – 1000S
Cs Cp 0.01pF – >1F
Ls Lp 0.01nH – 100KH
D 0.00001 – 9.9999
Q 0.1 – 9999.9
DCR 0.1mΩ – 100MΩ
Δ% -999.99% – 999.99%
Θ -180° – +180°

s, Delay Line, Filters, Resonators

  • Açıklama
    • Product Description:

      The 6370 LCR Meter that is provided with high accuracy and speed. It is a cost and multiple- function utility and is able to perform the component test entirely. The friendly operation interface design is in synchronization with the international trend and easy to operate to import into the production lines quickly. Because of the high-precision values of measured data and superior efficacy, it must be a necessary testing instrument of components and materials. It supports the interfaces of  RS232、Handler  and GPIB PC connections.

      • Basic accuracy up to 0.1%
      • LCR Meters are all compatible with convenient and efficient Meter Link Software
      • Overall Compact design, good for production line use
      • User-friendly interface, allows for production line personnel to operate efficiently

      Product Specification:

      Measurement Data

         
      Test Frequency 100Hz-20KHz
      Points Of Test Frequency 372 points
      Basic Accuracy ± 0.1%
      AC Drive Level 0.05V、0.1V、0.25V、0.5V、1V
      DC Drive Level 0.5V
      Output Impedance 100Ω
      Frequency trimming resolution 10 Digits
      Frequency output accuracy ± 0.01%
      AC test speed (Fast) 5 times/sec
      AC test speed (Medium) 3 times/sec
      AC test speed (Slow) 2 times/sec
      DC test speed (Fast) 5 times/sec
      DC test speed  (Medium) 3 times/sec
      DC test speed  (Slow) 2 times/sec
      Bias voltage 2V

      Specifications

         
      ESR Series/Parallel
      Mathematical Mode Absolute/Percentage
      Test Range Automatic/Lock
      Indicator Pass/Fail/Bias LED and Built-in speaker
      Output Terminal BNC connector
      Safety EN61010-1
      Immunity EN61326-1
      Emissions EN55022 Class A
      Calibration Open Circuit / Short Circuit: Avoid the misreading caused by the stray capacitance from kelvin clips and fixtures
      Power Supply Voltage 90-132Vac or 198-264Vac (selectable)
        Frequency 47-63Hz
      Display 240*64 dot-matrix display
      Interface RS-232, GPIB (option), Handler (option)
      Flash Memory Storage of 127 rewritable data sets
      Operation Auto、Manual、Remote Control、RS-232、GPIB (option)
      Environment Temperature : 10°C to 40°C    Humidity : 20 ~ 90%
      Dimension (W*H*D) 340x88x330 mm
      Weight 5 Kg

      Parameters Measurement Ranges

         
      |Z| R X 0.1 mΩ – 100MΩ
      |Y| G B 10nS – 1000S
      Cs Cp 0.01pF – >1F
      Ls Lp 0.01nH – 100KH
      D 0.00001 – 9.9999
      Q 0.1 – 9999.9
      DCR 0.1mΩ – 100MΩ
      Δ% -999.99% – 999.99%
      Θ -180° – +180°